Abstract
A cascaded H-bridge based pulse generator for transcranial magnetic stimulation is introduced. The system demonstrates complete flexibility for producing different shape, duration, direction, and rate of repetition of stimulus pulses within its electrical limits, and can emulate all commercial and research systems available to-date in this application space. An offline model predictive control algorithm, used to generate pulses and sequences, shows superior performance compared to conventional carrier-based pulse width modulation. A fully functioning laboratory prototype delivers up to 1.5 kV, 6 kA pulses, and is ready to be used as a research tool for the exploration of transcranial magnetic stimulation therapies by leveraging the many degrees-of-freedom offered by the design.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Conference proceedings. IEEE Applied Power Electronics Conference and Exposition
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.