Abstract
In manufacturing processes of semiconductor industry accurate detection and monitoring of water vapor concentration in trace amount is of great importance. The ability to perform reliable measurements in ultra-pure gases, with a wide dynamic range and low uncertainty, can have a substantial impact on product quality and process performances. Here, we report on the development of a second-generation comb-assisted cavity ring-down spectrometer and present H2O mole fraction measurements in high-purity N2 gas. Based on the use of a pair of phase-locked lasers and referenced to an optical frequency comb synthesizer, the spectrometer allowed to record high-quality absorption spectra in coincidence with the 32,2→22,1 H2O transition at 1.3946μm. Retrieval of water mole fractions, at levels as low as 380 part per billion, was accomplished through a careful spectra analysis procedure based on the use of refined line shape models which include speed-dependent effects. Measurements were performed with a statistical reproducibility of 5 parts per billion, for an integration time of about 0.2 s. The noise equivalent and minimum detectable absorption coefficients were found to be 3.1 × 10−11 cm−1/Hz and 6.5 × 10−12 cm−1, respectively. This latter allowed for a minimum detectable water mole fraction (limit of detection) of 160 parts per trillion. Finally, the main sources of systematic uncertainty have been discussed and quantified.