Abstract

We present a portable spectrometer that uses the frequency-stabilized cavity ring-down spectroscopy technique capable of high-precision measurements of trace water vapor concentration. Measuring one of the strongest rovibrational transitions in the ν1+ν3 water vapor combination band near ˜ν=7181.156 cm-1, we compare spectroscopic and thermodynamic determinations of trace water vapor in N2, and find systematic differences attributable to water vapor background effects and/or uncertainties in line intensities. We also compare the frequency-stabilized ring-down method with other cavity ring-down approaches that are based on unstabilized probe lasers and unstabilized ring-down cavities. We show that for the determination of water vapor concentration, the frequency-stabilized cavity ring-down method has the minimum measurement uncertainty of these techniques. The minimum noise-equivalent absorption coefficient of the spectrometer was 1.2×10-10 cm-1 Hz-1/2, which further corresponds to a minimum detectable water vapor mole fraction equal to 0.7×10-9 for an absorption spectrum of 10 minutes duration.