Abstract

A technique for volumetric blood flow measurement was developed by combining standard Doppler measurements with grey-scale decorrelation. Steered Doppler is used to determine the in-plane velocities, which are then used to extract the out-of-plane velocities from the temporal A-line decorrelation. As a result, a three-dimensional (3-D) vector flow field can be computed over the imaging plane using a single clinical transducer without knowledge of the vessel orientation. Volume flow is computed by integrating the out-of-plane flow over the vessel cross-section. The algorithm was tested using a scattering-enhanced fluid in a 6.4-mm diameter dialysis tubing. For a wide range of transducer angles, the volume flow was accurately measured to within 28% in these preliminary tests. (E-mail: tuthill@umich.edu)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.