Abstract

The Encore PV is a novel, real time 3D (i.e., 4D) ultrasound scanner capable of measuring blood flow. In vitro and in vivo volume flow measurements obtained from the Encore were compared to those from an invasive transit time flowmeter. A flow pump was used to simulate carotid type waveforms with pulsatile flow rates from 60 to 600 ml/min. Encore volume flows were compared to an invasive technique for volumetric blood flow measurement utilizing a transit time flowmeter. In vivo 10 s data sets of the volume flow in the distal aorta of 6 rabbits were obtained simultaneously with the Encore PV and the flowmeter. Mean and maximum volume flows were recorded with both units. Data was compared using linear regression and Bland-Altman analysis. It was found that volume flow measurements with this unit are comparable to those of an invasive flowmeter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.