Abstract

Beta Secondary Standard 2 (BSS 2) provides beta radiation fields with certified values of absorbed dose to tissue and the derived operational radiation protection quantities. As part of the quality assurance, the reliability of the CDTN BSS2 system was verified through measurements in the 90Sr/90Y and 85Kr beta radiation fields. Absorbed dose rates and their angular variation were measured with a 23392 model PTW extrapolation chamber and with Gafchromic radiochromic films on a PMMA slab phantom. The feasibility of using both methods was analyzed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call