Abstract

The Calibration Laboratory (LCI) at the Instituto de Pesquisas Energéticas e Nucleares (IPEN) is going to establish a Böhm extrapolation chamber as a primary standard system for the dosimetry and calibration of beta radiation sources and detectors. This chamber was already tested in beta radiation beams with an aluminized Mylar entrance window, and now, it was characterized with an original Hostaphan entrance window. A comparison between the results of the extrapolation chamber with the two entrance windows was performed. The results showed that this extrapolation chamber presents the same effectiveness in beta radiation fields as a primary standard system with both entrance windows, showing that any one of them may be utilized.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.