Abstract
The qualitative Roche HIV-1 DNA Amplicor assay has been used for the past 20 years to diagnose HIV infection in infants and young children but is being phased out; hence, alternative assays must be found. The Gen-Probe Aptima qualitative HIV-1 RNA assay is currently the only FDA-cleared HIV-1 nucleic acid assay approved for diagnosis, but data on the use of this assay with infant plasma are limited. We assessed Aptima's performance using control material for reproducibility and limit of detection and 394 plasma samples (0.2 to 0.5 ml) from HIV-exposed infected and uninfected infants and children for analytical sensitivity and specificity. Assays to assess within-run repeatability and between-run reproducibility indicated that the controls with 10,000 (5 of 5), 200 (5 of 5), 100 (16 of 16), 50 (12 of 12), and 25 (20 of 20) HIV-1 RNA copies/ml (cp/ml) were always positive, and negatives were always negative (20 of 20). The limit of detection was 14 cp/ml, as determined by probit analysis. The analytic sensitivity of the assay was 99.5% (189/190 samples; 95% confidence interval [CI], 97.1 to 99.9%) and specificity was 99.5% (199/200 samples; 95% CI, 97.2 to 99.9%). These results suggest that the assay is suitable for early infant diagnosis of HIV-1.