Abstract

A new procedure including high performance liquid chromatography in combination with photothermal interference spectroscopy as detection device (HPLC/PIS) has been proposed, optimized and its figures of merit for pesticide residue analysis are shown. The flowing sample under study is set in one arm of a Mach-Zehnder interferometer, and its refractive index is modulated by a periodically chopped continuous wave argon ion laser. As chopper, an acousto optical modulator has been introduced to switch the excitation laser beam between different lines (457 nm, 488 nm, 514 nm) simultaneously. Thus a multi component analysis can be realized either by using an HPLC-system in front of the PIS device or by a multi line Ar(+)-laser, directly. The limit of detection of the HPLC/PIS system reached 71 microg/l of the pesticide di-nitro-ortho-cresol (DNOC).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.