Abstract
The accurate and timely detection of biochemical coagulation indicators is pivotal in pulmonary and critical care medicine. Despite their reliability, traditional laboratories often lag in terms of rapid diagnosis. Point-of-care testing (POCT) has emerged as a promising alternative, which is awaiting rigorous validation. We assessed 226 samples from patients at the First Affiliated Hospital of Guangzhou Medical University using a Beckman Coulter AU5821 and a PUSHKANG POCT Biochemistry Analyzer MS100. Furthermore, 350 samples were evaluated with a Stago coagulation analyzer STAR MAX and a PUSHKANG POCT Coagulation Analyzer MC100. Metrics included thirteen biochemical indexes, such as albumin, and five coagulation indices, such as prothrombin time. Comparisons were drawn against the PUSHKANG POCT analyzer. Bland-Altman plots (MS100: 0.8206‒0.9995; MC100: 0.8318‒0.9911) evinced significant consistency between methodologies. Spearman correlation pinpointed a potent linear association between conventional devices and the PUSHKANG POCT analyzer, further underscored by a robust correlation coefficient (MS100: 0.713‒0.949; MC100: 0.593‒0.950). The PUSHKANG POCT was validated as a dependable tool for serum and whole blood biochemical and coagulation diagnostics. This emphasizes its prospective clinical efficacy, offering clinicians a swift diagnostic tool and heralding a new era of enhanced patient care outcomes.
Published Version