Abstract

External hip protectors are used by the elderly in preventing hip fracture due to sideway falls. There are some commercial hip protectors which has both energy absorbing and energy shunting properties. In this study, a novel hip protector using shear thickening polymer (STP) is studied. The purpose of this work is to determine the optimal thickness of STP needed for maximum force attenuation. A mechanical test rig to simulate a person falling with sufficient impact energy to fracture the greater trochanter if unprotected was used together with biofidelic femur model which simulates the layer of flesh with skin. 8mm of STP together with 5mm foam gives the best force attenuation. When comparing the overall thickness with commercial hip protectors, STP hip protectors tested have much less thickness. Reduced thickness increases the compliance and comfort of STP hip protectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.