Abstract

The purpose of this investigation was to assess the reliability of the i-STAT clinical portable analyser and CG(4)(+) cartridge measures of blood pH, partial pressures of O(2) (pO(2)) and CO(2) (pCO(2)), bicarbonate ([HCO(3)(-)]), base excess (BE), total carbon dioxide (TCO(2)), oxygen saturation (sO(2)) and blood lactate ([BLa(-)]) at various exercise intensities. A comparison between [BLa(-)] measured by the i-STAT and the Accusport lactate analysers during an intermittent treadmill run was also undertaken. The technical error of measurement (TEM%) at rest, at moderate (200W) and maximal exercise (V O(2)max) was acceptable (<15%) for all parameters. The intra-class correlation coefficients for each analyte ranged from weak-to-strong across resting (ICC=0.19-0.96) and moderate (ICC=0.30-0.96) exercise intensities. The ICC for all analytes were observed to be strong following maximal intensity exercise (ICC=0.77-0.95). The comparison of the [BLa(-)] measures between the i-STAT and Accusport showed that the difference between measures was acceptable at both low (<4mmolL(-1)) (-0.39+/-0.27mmolL(-1)), moderate to high concentrations (>4mmolL(-1)) (0.58+/-1.22mmolL(-1)), and across all [BLa(-)] data (0.36+/-1.13mmolL(-1)). In conclusion, the i-STAT clinical analyser and CG(4)(+) cartridge provides reliable measures of a number of blood parameters across exercise intensities. The [BLa(-)] measures from the i-STAT analyser are consistent with that of the Accusport lactate analyser.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call