Abstract

Measurements have been made of the tap densities and angles of internal flow of physically mixed and formulated mixtures of oxytetracycline and lactose powders. The differences between the two types of mixtures are ascribed to differences in the packing and surface morphologies of the particles.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.