Abstract

The purpose of this study was to evaluate the stress distribution on the temporomandibular joint (TMJ) prosthesis and contralateral natural TMJ with finite-element analysis. A TMJ implant was used to create a 3-dimensional model with a computer. This model was integrated with a mandible model created with a computer by use of computed tomography images, similar to a real TMJ replacement procedure. Masticatory loads were applied to the model. The loads transmitted to the TMJ prosthesis and contralateral healthy joint were evaluated by means of finite-element analysis. In the model without the TMJ prosthesis, maximum von Mises stress was 252.697N/mm2 at the condyle and 5.418N/mm2 at the disc. In the model with the unilateral TMJ prosthesis, maximum stress at the joint prosthesis was 792.681N/mm2. In the contralateral natural joint, maximum stress was 268.908N/mm2 at the condyle and 8.357N/mm2 at the disc. In the TMJ model with the unilateral total TMJ prosthesis, increased stress values were observed at the disc and condyle of the contralateral natural TMJ.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.