Abstract
A digital multimeter automatic calibration system is developed by modifying the traditional multifunctional calibrator, designing the controller and application program. PID closed-loop feedback control algorithm is used to slove the problem of low accuracy of digital multimeter calibration with a small range of 200mV. The test results show that the system has good measurement stability and can realize automatic calibration of digital multimeter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.