Abstract

The calibration parameters of a high resolution two-chamber alpha spectrometer dedicated to detect alpha radiation in environmental samples were evaluated and confirmed using a certified standard source containing 237Np, 241Am and 244Cm nuclides. The resolution and efficiency characteristics for the two ion-implanted solid state detectors (ULTRA AS series) in this system were evaluated at all possible detector-source distances. The characteristics related to the performance of alpha detectors against the alpha radiation emitted from the standard source used in the measurements and the results obtained were compared in detail.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.