Abstract
Studies and evaluations of the main parameters of a high resolution multi-chamber alpha spectrometer and its full calibration using standard sources of 241Am, 233U and 244Cm are presented for the validation of the complex method of characterization of an alpha spectrometer dedicated for alpha radioactive concentration measurements of various environmental samples. The resolution and efficiency were determined at all possible source-detector distances for all eight solid state ORTEC ULTRA-AS detectors. In addition, the solid angles associated to different measurement geometries and the repeatability of the results were assessed. A detailed inter-comparison of the results was performed drawing important conclusions regarding the quality of the alpha detector response to the alpha radiation of the standard sources used in measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.