Abstract

The Fourier transform ion cyclotron resonance method holds the lead in mass accuracy and resolving power among all other mass spectrometry methods. The dynamically harmonized cell is largely responsible for the supremacy. This cell has an ideal hyperbolic trapping potential after averaging over fast cyclotron motion. Recently we have introduced an open modification of the cell (especially useful with ultrahigh magnetic fields) and have found the analytical solution for the averaged potential inside it. The voltage on specific "regularizing" electrodes determines how close a potential is to the hyperbolic one. In this article, we find the optimal voltage on these "regularizing" electrodes analytically. This will assist with both further analysis and tuning of the trap after manufacturing.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.