Abstract

Experimental data on the temperature behavior of quartz crystal microbalances (QCMs) rigidly coated with solid polymer films of hexamethyldissiloxane (HMDSO) for gas and liquid-phase sensor applications are provided. 16MHz AT-cut quartz resonators used as QCMs are coated at seven different HMDSO thicknesses in the 70–400nm range by using a RF-plasma polymerization process. Their quasi cubic temperature frequency characteristics (TFCs) are measured over a temperature of −50°C to 90°C for each film thickness. A cubic spline approximation of the experimental data reveals a periodic oscillating behavior of the temperature coefficients of the TFC, while its inflection point oscillates between 24°C and 28°C in the above thickness range. Devices coated at 150, 260 and 380nm, i.e. at a thickness period of 110–120nm, were found to have identical thermal stability and TFC behavior. A simple practical method for correcting the sensor readings for thermally induced frequency shifts, based on an additional temperature measurement at each sensor reading, is suggested and verified experimentally.