Abstract
The relationships among frequency changes on a film-coated quartz crystal microbalance, thickness changes, and dc resistance changes have been investigated for carbon black-insulating polymer composite vapor detectors. Quartz crystal microbalance (QCM) measurements and ellipsometry measurements have been performed simultaneously on polymer films that do not contain carbon black filler to relate the QCM frequency change and the ellipsometrically determined thickness change to the analyte concentration in the vapor phase. In addition, quartz crystal microbalance measurements and dc resistance measurements on carbon black composites of these same polymers have been performed simultaneously to relate the QCM frequency change and dc electrical resistance response to the analyte concentration in the vapor phase. The data indicate that the dc resistance change is directly relatable to the thickness change of the polymers and that a variety of analytes that produce a given thickness change produce a constant resistance change for each member of the test set of polymers investigated in this work.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Analytical Chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.