Abstract

Extracorporeal cytokine removal may be desirable. We sought to assess extracorporeal blood purification (EBP) techniques for cytokine removal in experimental animal studies. We conducted a targeted, systematic search and identified 17 articles. We analyzed cytokine clearance, sieving coefficient (SC), ultrafiltrate (UF) concentration, and percentage removal. As this review concerns technical appraisal of EBP techniques, we made no attempts to appraise the methodology of the studies included. Results are in descriptive terms only. Applying predicted clearance for 80 kg human, high volume hemofiltration (HVHF) techniques and plasmafiltration (PF) showed the highest rates of cytokine removal. High cutoff (HCO)/HF and PF techniques showed modest ability to clear cytokines using low to medium flows. Standard hemofiltration had little efficacy. At higher flows, HCO/HF achieved clearances between 30 and 70 ml/min for IL-6 and IL-10. There was essentially no removal of tumor necrosis factor (TNF)-alpha outside of PF. Experimental animal studies indicate that HVHF (especially with HCO filters) and plasmafiltration have the potential to achieve appreciable IL-6 and IL-10 clearances. However, only PF can remove TNF-alpha reliably.