Abstract

Thrombosis within extracorporeal membrane oxygenation (ECMO) circuits is a common complication that dominates clinical management of patients receiving mechanical circulatory support. Prior studies have identified that over 80% of circuit thrombosis can be attributed to tubing-connector junctions. A novel connector was designed that reduces local regions of flow stagnation at the tubing-connector junction to eliminate a primary source of ECMO circuit thrombi. To compare clotting between the novel connectors and the traditional connectors, both in vitro loops and an in vivo caprine model of long-term (48 h) ECMO were used to generate tubing-connector junction clots. In vitro, the traditional connectors uniformly (9/9) formed large thrombi, while novel connectors formed a small thrombus in only one of nine (p < 0.0001). In the long-term goat ECMO circuits, the traditional connectors exhibited more thrombi (p < 0.04), and these thrombi were more likely to protrude into the lumen of the tubing (p < 0.001). Both in vitro and in vivo validation experiments successfully recreated circuit thrombosis and demonstrate that the adoption of novel connectors can reduce the burden of circuit thrombosis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.