Abstract

We propose methods to suppress the interference noise (IN) of an all-fiber white-light interferometer (WLI), which is caused by the residual Fresnel reflective beam. The methods are proposed to ensure a wide dynamic range and enhance the accuracy of measurement. IN can cause misjudgment of the realistic optical characteristic parameters, such as the fault diagnosis and polarization coupling of the optical devices under test. In addition, IN can reduce the dynamic range of the WLI. The optical path mismatch (OPM) method and the intensity suppression (IS) method by changing the positions and intensity of the IN are presented. The two suppression methods can successfully restrain IN caused by the Fresnel reflection beams. The experimental results show that IN is successfully suppressed by the OPM method or restrained by IS method in interferograms, and the dynamic range can achieve 85dB without IN.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.