Abstract

AbstractPurposeIn our study basic dosimetric properties of a flattening filter free 6 MV photon beam shaped by multileaf collimators (MLC) is examined using the Monte Carlo (MC) method.Methods and MaterialsBEAMnrc code was used to make a MC simulation model for 6 MV photon beam based on Varian Clinic 600 unique performance linac, operated with and without a flattening filter in beam line. Dosimetric features including central axis depth dose, beam profiles, photon and electron spectra were calculated and compared for flattened and unflattened cases.ResultsDosimetric field size and penumbra were found to be smaller for unflattened beam, and the decrease in field size was less for MLC shaped in comparison with jaw-shaped unflattened beam. Increase in dose rate of >2·4 times was observed for unflattened beam indicating a shorter beam delivery time for treatment. MLC leakage was found to decrease significantly when the flattening filter was removed from the beam line. The total scatter factor showed slower deviation with field sizes for unflattened beam indicating a reduced head scatter.ConclusionsOur study demonstrated that improved accelerator characteristics can be achieved by removing flattening filter from beam line.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.