Abstract
The energy response calibration of the commonly used high pressure ionization chamber is very difficult to obtain when the gamma-ray energy is more than 3 MeV. In order to get the calibration of the higher part of the high pressure ionization chamber, we use the Fluka Monte Carlo program to perform the energy response in both the spherical and the cylindrical high pressure ionization chamber which are full of argon gas. The results compared with prior study when the gamma-ray energy is less than 1.25 MeV. Our result of Monte Carlo calculation shows agreement with those obtained by measurement within the uncertainty of the respective methods. The calculation of this study is significant for the high pressure ionization chamber to measure the high energy gamma-ray.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.