Abstract
A detector from single crystal synthetic diamond with conducting wires has been prepared with an improved femto-second laser process. The detector was characterised with a 4.5MeV proton micro-beam (Ruđer Bos̆ković Institute, Zagreb). The charge collection efficiency and the transient current response have been investigated with high spatial resolution. A hexagonal and square cell geometry is investigated. Both cell geometries show full charge collection at 40V bias voltage, and little charge sharing between neighbouring cells. The experimental data is compared to a simulation and qualitative agreement is observed.
Submitted Version (Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have