Abstract
The fundamental properties of the laminar flow torch (LFT) designed in our laboratory are described. Noise power spectra are presented which show that the principle source of noise in a conventional tangential flow torch (TFT) has been removed in the laminar flow torch. Spatial profiles of Ca atom and ion emission intensities are observed to be displaced towards the induction coil in comparison with those in a conventional tangential flow torch. Electron densities measured in the LFT are lower than those in the TFT but Ca ion to atom ratios are higher in the LFT and hence ionisation temperatures are comparable to those in the TFT. Detection limits in the LFT are therefore superior by an order of magnitude compared with the TFT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.