Abstract
Noise power spectra of emission signals from the laminar flow torch (LFT) designed in our laboratory and a conventional tangential flow torch (TFT) are presented and described. It is shown that the fundamental frequency of rotation in a conventional torch is completely removed in the laminar flow torch. Moreover, the use of an extended torch is essential in removing air entrainment effects and also for the maintenance of undisturbed laminar flow. The interference effects of phosphate and sodium on calcium emission are assessed in the laminar flow torch and compared with those in a conventional torch. Phosphate interference is not observed in the laminar flow torch while the effect of sodium on calcium emission is shown to be of the same magnitude in both torches.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.