Abstract
The use of the microwave-induced plasma with a new low-flow plasma centering torch has been evaluated. Data were collected through a computer-controlled, background-correcting polychromator system with two versions of the newly developed laminar flow torch. It showed sensitivities slightly better than the 0.5-mm-i.d. capillary quartz open tube torch. Detection limits were between 8 and 60 pg/s for carbon, hydrogen, bromine, chlorine, and fluorine—a considerably improved performance over the tangential flow torch. Torch lifetime and plasma stability are comparable to or better than the high-flow tangential flow torch with three-hour and three-day RSDs of 2% and 6%. Centered plasmas may be sustained with flow rates as low as 5 mL/min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.