Abstract

Pulse wave velocity (PWV) has become a gold standard index to quantify the stiffness of the aorta and is a predictor of cardiovascular events. A recent paper compared the pOpmètreⓇ, a device for measuring the finger-toe PWV, with other techniques and demonstrated its accuracy and validity. However, human devices do not allow the advancement of our knowledge on conditioning mechanisms. Based on its human validation, a new device, pOpetⓇ 1.0 system was designed for estimation of PWV in small animals and this present study aimed to standardize the pOpetⓇ 1.0 for estimation of arterial stiffness in rats, and to confirm its liability and stability as well as the reproducibility of assessments. Therefore several precautions were taken into consideration like as the correct position of the animal and photodiodes according to manufacturers’ suggestions. Results indicated that estimation of PWV through the new pOpetⓇ 1.0 device exhibits good internal consistency, stability and objectivity in all tests performed between days and evaluators. Importantly, data suggest for the first time that this new device is able to detect changes in arterial stiffness that are conditioned by age and pressure-related arterial remodeling.• This new pOpetⓇ device is able to detect changes in vessel structure.• This new pOpetⓇ device exhibits good internal consistency, stability and objectivity in all tests performed• Correct position of the animal and photodiodes are crucial to obtain a very stable signal.

Highlights

  • ObjectivesIn this study we aimed to standardize the pOpet R for estimation of arterial stiffness in rats, and to confirm its liability and stability as well as the reproducibility of assessments

  • Pulse wave velocity (PWV) has become a gold standard index to quantify the stiffness of the aorta and is a predictor of cardiovascular events

  • 6 male Wistar rats (~300 g) were used. Since this device was commercialized based on its humans study, we provide in this article some important information about correct position of the animal and photodiodes followed by several precautions, which are crucial to obtain a very stable signal

Read more

Summary

Objectives

In this study we aimed to standardize the pOpet R for estimation of arterial stiffness in rats, and to confirm its liability and stability as well as the reproducibility of assessments. Study 2: This study aimed to correlate the estimated value of PWV by pOpet R 1.0 with ageinduced changes in vessels morphology

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.