Abstract

The apparatus function of prism or grating spectroscopes with finite slits is derived for the general case of partially coherent illumination of the slit. Numerical results are presented for the border cases of coherent and incoherent slit illumination, as well as for full and partial illumination of the spectroscope aperture.The dependence of line intensity and line width on slit width, as well as the optimum slit width, are determined for each of these cases. It is shown that the performance of the spectroscope reaches an optimum when the aperture is at least fully illuminated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.