Abstract

The authors critically analyze the paper “Corrections for matrix effects in X-ray fluorescence analysis” [1], which was declared by R. Rousseau as a tutorial but contains some inaccuracies. Conclusions are confirmed by the results of experiments and theoretical calculations. It is shown that uniform linear calibration function Ii/Iim = f(Rith) can hardly be obtained for test samples of any composition because of the difference in their microstructure and the variation of the background. At the same time, in the XRF analysis of multicomponent materials, such as steels, the Lucas-Tooth and Jongh equations provide a good alternative to the Rousseau algorithm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.