Abstract

A new cobalt ions pre-concentration method, optimised by fractional factorial design, using multiwall carbon nanotubes (MWCNTs) with further Graphite Furnace Atomic Absorption Spectrometry (GFAAS) quantification is described. The method explores the high chemical and physical stability of MWCNTs for improving the detectability of GFAAS. It is based on off-line pre-concentration of 20.0 mL of sample previously buffered (pH 8.82) on MWCNTs at a flow rate of 10.0 mL min−1. After the pre-concentration procedure, the elution step was carried out with 500 µL of 0.524 mol L−1 HNO3 solution at a flow rate of 2.0 mL min−1. Fractional factorial designs and response surface methodology were employed for optimisation of all chemical parameters involved in the pre-concentration procedure, including pre-concentration flow rate, buffer and eluent concentration, sample pH and elution volume. The method provides a linear calibration range from 0.03 up to 7.00 µg L−1 with linear correlation coefficient higher than 0.9994 and limits of detection and quantification of 0.01 and 0.03 µg L−1, respectively. Repeatability of the six measurements was found to be 2.38 and 1.84% for 0.3 and 4.5 µg L−1 cobalt concentration, respectively. By pre-concentrating 20.0 mL of sample, a pre-concentration factor (PF) of 19.10-fold and a consumption index of 1.05 mL were obtained. The pre-concentration efficiency (PE) was found to be 9.55 min−1. The proposed method was successfully applied for the pre-concentration and determination of cobalt in water and urine samples with satisfactory recovery values.