A method for determination of cobalt in water by flow injection spectrophotometry based on the catalytic effects of cobalt on the oxidation reaction of arsenazo III by potassium bichromate in HAc-NaAc buffer solution of pH 4.6 was established. The relationship of the reduced absorbency to the concentration of cobalt was good linearity and the maximum absorption peak located at 550 nm. The affecting factors on the strength of absorption including acidity, dosage, temperature, coexisting ions and so on, and the optimal conditions of the reaction in the system were investigated. Calibration graphs were linear for ranges of 0 to 20 μg/mL and a detection limit of 0.3 μg/mL cobalt was achieved. The proposed method was successfully applied to the analyses of industrial waste water samples, with relative standard deviation of less than 1.5% and the recoveries are 95.1% to 96.8%for the determination of cobalt.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.