Abstract
A practicable method was developed for the routine determination of formaldehyde in air. Formaldehyde is sampled in small sampling tubes filled with Chromosorb P, coated with 2,4-dinitrophenylhydrazine (= DNPH). The sorbent is extracted with acetonitrile, and the different DNPhydrazones were separated by reversed-phase HPLC. UV-detection at γ = 345 nm was compared with electrochemical detection. The detection limit was 100 pg for UV- and 50 pg for electrochemical detection with a linear range of more than 3 decades, respectively. For the determination of formaldehyde in air the detection limit is 10 ppb with an average recovery of 99.3% and an estimated relative standard deviation S of 1.5%. This method was compared with the sampling by impingers using the same separation and detection method.