Abstract

A new soft x-ray pulse height analyzer (PHA), based on a Silicon Drift Detector (SDD) linear array consisting of 15 SDD detectors, has been installed in the HT-7 tokamak. The energy resolution of 150–180 eV at photon energy of 6 keV is achieved for the SDD with Peltier cooling. The effective time response of the SDD PHA is 50 ms. The profiles of electron temperature and the intensity of metallic impurities can be obtained with a spatial resolution of 3 cm. The performance and first experimental results from the new PHA system are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.