Abstract
Precision measurement of silicon wafer resistivity has been using single-configuration Four-Point Probe(FPP) method. This FPP method have to applying sample size, shape and thickness correction factor for a probe pin spacing to precision measurement of silicon wafer. The deference for resistivity measurement values applied correction factor and not applied correction factor was about 1.0 % deviation. The sample size, shape and thickness correction factor for a probe pin spacing have an effects on precision measurement for resistivity of silicon wafer.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have