Abstract
In the present paper the use of Prussian blue film modified-palladized aluminum electrode (PB/Pd-Al) prepared by a simple and rapid electroless method for the determination of codeine (CO) and morphine (MO) is described. In the first part of the work the oxidation of CO and MO at the modified electrode was investigated by cyclic voltammetry. The electro-oxidation pathway for the both two compounds in the neutral medium (pH 6) was suggested. In optimum conditions the hydrodynamic amperometry was used for the determination of CO and MO at the μM concentration level. It is found that the calibration graphs of CO and MO are linear in the concentration ranges 2–30 and 2–50 μM, respectively. The detection limit of the method was about 0.8 μM. The possibility of using differential pulse voltammetry as well as hydrodynamic amperometry for simultaneous determination of CO and MO at a mixture solution is demonstrated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.