Abstract
A new method for the simultaneous determination of the divalent ions of cadmium, cobalt, copper, lead, manganese, nickel and zinc in binary, ternary and quaternary mixtures based on the application of a multiple linear regression programme to the spectra of the complexes of these ions with 4-(pyridyl-2-azo)resorcinol (PAR) was developed. The detection limits afforded by the proposed method range from 0.02 μg/ml for Co(II), Mn(II), Ni(II) and Zn(II) to 0.10 μg/ml for Pb(II). The method is rapid and quite simple as it entails no prior separation, and was successfully applied to the analysis for the above-mentioned ions in synthetic samples and real alloys.