Abstract

A sensitive and rapid ultra performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method was developed to determine olmesartan and amlodipine levels in human plasma and urine simultaneously. Chromatographic separation was carried out on an Acquity UPLC BEH C18 column and mass spectrometric analysis was performed using a QTrap5500 mass spectrometer coupled with an electro-spray ionization (ESI) source in the positive ion mode. The MRM transitions of m/z 447→207 and 409→238 were used to quantify olmesartan and amlodipine, respectively. This assay method has been fully validated in terms of selectivity, linearity, lower limit of quantification (LLOQ), accuracy, precision, stability, matrix effect and recovery. The linearity of this method was found to be within the concentration range of 0.2-500ng/mL and 4-5000ng/mL for olmesartan in human plasma and urine and 0.1-50ng/mL and 2-1000ng/mL for amlodipine in human plasma and urine. Only 2min were needed for an analytical run. This assay was used to support a clinical study where multiple oral doses were administered to healthy Chinese subjects to investigate the pharmacokinetics of olmesartan and amlodipine.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.