A sensitive and convenient high performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS) method was developed to determine levophencynonate and demethyl levophencynonate levels in human plasma simultaneously. Chromatographic separation was achieved on a SHIMADZU Shim-Pack XR C8 column and mass spectrometric analysis was performed by an API5000 mass spectrometer coupled with an electro-spray ionization (ESI) source in the positive ion mode. The MRM transitions of m/z 358.4→156.4 and 344.5→144.2 were used to quantify levophencynonate and demethyl levophencynonate, respectively. This analytical method was fully validated with specificity, linearity, lower limit of quantitation (LLOQ), accuracy, precision, stability, matrix effect and recovery. The linearity of this method were developed to be within the concentration ranges of 10-4000pg/mL for levophencynonate and 25-8000pg/mL for demethyl levophencynonate in human plasma. This method was used in a clinical study which was administrated with single oral dose for Chinese healthy subjects to investigate the pharmacokinetics of levophencynonate and demethyl levophencynonate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.