Abstract

In the present paper, the simultaneous determination of codeine (CO) and caffeine (CF) is described by the use of single-walled carbon nanotubes modified carbon-ceramic electrode (SWCNT/CCE); prepared via a simple and rapid method. The results show that the SWCNT/CCE exhibits excellent electrochemical catalytic activity toward the oxidation of these compounds with respect to the bare CCE and offers two anodic peaks at 1.05 and 1.38V vs. saturated calomel electrode for oxidation of CO and CF, respectively. Differential pulse voltammetry was used for simultaneous determination of CO and CF at micromolar concentration level. In the optimum conditions, it is found that the calibration graphs for CO and CF are linear in the concentration ranges 0.2–230 and 0.4–300μM with detection limits of 0.11 and 0.25μM for CO and CF, respectively. The SWCNT/CCE presents good stability, reproducibility, and repeatability and the proposed method has been successfully applied for determination of CO and CF in some pharmaceutical, drinking and biological samples with high recovery rate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.