Abstract
The effect of solvent composition on the uv–visible spectrum of methyl orange was investigated for application to the quantitative determination of ethanol. At fixed pH, there was a hypsochromic shift of the absorbance band of methyl orange with increasing ethanol concentration. Using acetate buffer at pH 3.40 the change of absorbance at 530 nm of a solution of methyl orange containing known concentrations of standard ethanol was measured to provide a calibration curve. In order to apply this method to the analysis of alcoholic samples, such as distilled spirits, blended spirits and liquid herbal medicines, a simple gas-diffusion unit coupled with flow system was employed to separate the ethanol from sample matrices. Using the gas diffusion-flow system and employing an evaporation time of 2 min, a linear calibration range of 5–45% (v/v) ethanol was achieved ((ΔA = (0.0078 ± 0.0002)x(ethanol, %(v/v)) + (0.040 ± 0.005)), r2 = 0.998). The limit of detection (3σ blank/slope) was 2.23% (v/v). The developed gas diffusion-flow system was applied to the analysis of colorless distilled spirits, yellow blended spirits and dark brown herbal medicines that are available in the local markets of Bangkok, Thailand. Validation of the method was carried out by comparing the results with analysis using gas chromatography. There was no statistically significant difference at the 95% confidence level for all alcoholic samples analyzed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.