Abstract
This work presents new design of a gas diffusion unit, called ‘membraneless gas diffusion (MGD) unit’, which, unlike a conventional gas diffusion (GD) unit, allows selective detection of volatile compounds to be made without the need of a hydrophobic membrane. A flow injection method was developed employing the MGD unit to determine ethanol in alcoholic drinks based on the reduction of dichromate by ethanol vapor. Results clearly demonstrated that the MGD unit was suitable for determination of ethanol in beer, wine and distilled liquors. Detection limit (3S/N) of MGD unit was lower than the GD unit (GD: 0.68%, v/v; MGD: 0.27%, v/v). The MGD design makes the system more sensitive as mass transfer is more efficient than that of GD and thus, MGD can perfectly replace membrane-based designs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.