Abstract

To improve signal-to-noise ratio (SNR) of intermolecular double-quantum coherence (iDQC) MRS on a 3 Tesla (T) whole-body scanner. A 32-channel phased array coil was used to acquire iDQC signal of a MRS phantom in the presence of large field inhomogeneity. The obtained individual spectra from the array elements were combined together in the time domain using a multichannel nonparametric singular value decomposition algorithm. The results were compared quantitatively with those acquired with a circularly polarized (CP) head coil. The achieved gain in SNR ranges from 1.63 to 2.10 relative to the CP coil, mainly depending on the relative position between the surface of the phased array coil and the voxel of acquisition. SNR enhancement of iDQC MRS in inhomogeneous fields on a 3T whole-body scanner is feasible with phased array coils. This can facilitate iDQC applications of high-resolution in vivo spectroscopy in the presence of field inhomogeneity for potential disease diagnosis in humans.