Abstract

The signal-to-noise ratio (SNR) and parallel imaging (PI) performance of two commercial phased-array coils (PACs) were examined in magnetic resonance imaging (MRI) of the brain. All measurements were performed on a 3.0T MRI instrument. The SNR and PI performance were evaluated with 32-channel and 15-channel PACs. A gradient echo sequence was used for obtaining images of a phantom. SNR and geometry factor (g-factor) maps were calculated from two images with identical parameters. Horizontal and vertical profiles were taken through the SNR maps in the axial plane. The average g-factor was measured in a circular region of interest in the g-factor maps for the axial plane. The SNR map of the 32-channel coil showed a higher SNR than that of the 15-channel coil at the phantom's posterior and lateral surfaces. The SNR profiles for the 32-channel coil also showed a 1.3-fold increase at the phantom's center. The average g-factor of the 32-channel coil was lower than that of the 15-channel coil at the same acceleration factor. These results indicate that the 32-channel coil can provide a higher spatial resolution and/or a faster imaging speed. Horizontal and vertical profiles are useful for evaluation of the performance of commercially available PACs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.