Abstract

A simple method is described for separation of interference by co-existing methyl chloride and acetaldehyde in gas chromatographic quantitation of low levels of chloroethylene in air. Methyl chloride and acetaldehyde were successfully separated from chloroethylene by FID-gas chromatography using a column of 5.5% Ucon oil LB 550 X and Porapak QS. Relative deviation of peak area was found less than 5% on reproducibility of quantitative analysis, and the limit of detection was 1-4 ppb of chloroethylene.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.