Abstract

Analytical conditions of gas chromatographic separation of triglycerides were described. On-column injection system using glass column was superior in higher trighycerides to the conventional sampling method. It was demonstrated that OV-17 has excellent stability up to temperature 400° comparing with other liquid phases SE-30, OV-1 and Versamid 900 by the thermal balance analysis. Reproducibility of quantitative analysis on triglycerides was studied by square method. Relative deviation of each peak area was found less than 5%. Under the beat condition, triglycerides of milk or its products can be determined using ratio index of lower to higher ones.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.