Abstract
The Sample Oxidizer automatically prepares samples for liquid scintillation counting containing singlelabel tritium (^3H), single-label carbon-14 (^<14>C) or both nuclides in double-label. Double-label samples of tritium and carbon-14 are chemically separated during preparation. Recovery of tritium was about 90% with 0.2 gram of most materials. Recovery of carbon-14 was about 85%. Memory effect for tritium and carbon-14 was less than 0.6% with most samples. But Sample Oxidizer was not suitable for fatty samples. We adopt the "Sample Oxidizer Method" as to tissues, the "Hyamine 10X Method" as to fatty samples and the "Direct Method" as to serum.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.