Abstract

The objective of this study is to evaluate the safety and efficacy of a new transcutaneous bone-conduction implant (BCI BB) in patients with conductive and mixed hearing loss or with single-sided deafness (SSD), 1year after surgical implantation. The study design is multicentric prospective, intra-subject measurements. Each subject is his/her own control. The setting is nine university hospitals: 7 French and 2 Belgian. Sixteen subjects with conductive or mixed hearing loss with bone-conduction hearing thresholds under the upper limit of 45dB HL for each frequency from 500 to 4000Hz, and 12 subjects with SSD (contralateral hearing within normal range) were enrolled in the study. All subjects were older than 18years. The intervention is rehabilitative. The main outcome measure is the evaluation of skin safety, audiological measurements, benefit, and satisfaction questionnaires with a 1-year follow up. Skin safety was rated as good or very good. For the mixed or conductive hearing loss groups, the average functional gain (at 500Hz, 1, 2, 4kHz) was 26.1dB HL (SD 13.7), and mean percentage of speech recognition in quiet at 65dB was 95% (vs 74% unaided). In 5/6 SSD subjects, values of SRT in noise were lower with BB. Questionnaires revealed patient benefit and satisfaction. The transcutaneous BCI is very well tolerated at 1-year follow up, improves audiometric thresholds and intelligibility for speech in quiet and noise, and gives satisfaction to both patients with mixed and conductive hearing loss and patients with SSD.